The test task
SMD capacitors are fed automatically from a tray to the test station. Each capacitor must pass a high-voltage test in which the leakage current is measured. Testing parts one by one would make the hipot station the bottleneck of the line.
Test conditions used in this application
| Parameter | Setting |
|---|---|
| Device under test | SMD capacitors, automated feeding |
| Test | Withstand voltage with leakage current measurement |
| Test voltage | 4.5 kV |
| Parts per test cycle | Several capacitors in parallel |
| Result handling | Pass/fail per channel, sent to the automation system |
Why a parallel multi-channel hipot tester
- One instrument, up to eight parallel channels. A TH9010 tests up to eight channels at the same time and gives a separate pass/fail result for each channel.
- Expandable. With external four-channel scanners, one unit can be extended to up to 128 test points.
- Less space on the line. One parallel tester replaces several single-channel testers.
- Automation interfaces. Communication and handler interfaces allow full integration into the line control.
How it works
- The feeder places a group of capacitors in the multi-probe fixture.
- The tester applies the test voltage to all channels at the same time.
- Leakage current is measured on each channel and compared with its limit.
- Pass/fail results per channel go to the line control, which sorts the parts.
Before you plan a similar set-up
- The charging current of capacitors can trigger a leakage alarm at the start of the test. Set the ramp time and judgement delay to suit the capacitance of the part.
- Use high-voltage cables and fixtures rated for the test voltage, and provide a safety enclosure or light curtain on the line.
- Check with us how many channels you need per cycle before choosing between one tester with scanners and several testers.
Frequently asked questions
How many parts can the TH9010 test at the same time? Up to eight channels in parallel per unit, with a separate result for each channel.
Can the TH9010 be connected to automation equipment? Yes. In this application it runs on an automated line with tray feeding.