Application example

Inductor DC-bias (saturation) and inter-turn testing

An LCR meter with a DC bias current source measures inductance and Q under load current; an impulse winding tester adds a check for shorted turns. Two set-ups for 15 A and 40 A.

Based on a customer project. The customer's name is not published. Test conditions are the customer's, not instrument limits.

The test task

As the DC current increases, the core of an inductor approaches saturation and the inductance drops. A plain LCR measurement at zero bias does not show this. Manufacturers and developers of magnetic components therefore measure inductance and Q at the rated DC current. Production test often adds an inter-turn (layer-to-layer) insulation test to find shorted windings.

Example A: ferrite-core inductors at up to 15 A

A developer and manufacturer of magnetic materials and components already used an LCR meter without a bias source. It needed to see how inductance and Q fall when a DC bias of up to 15 A is applied.

ItemRole
TH2838 precision LCR meterMeasures L and Q
TH1779A DC bias current sourceSupplies the DC bias current

Example B: inductors and transformers at 40 A, with inter-turn test

A manufacturer of chargers, adapters and power transformers needed to test inductors and transformers with a saturation current of 40 A, and to detect shorts between winding layers.

ItemQtyRole
TH2840B / TH2840BL LCR meter1Measures inductance under bias
TH1779A DC bias current source1Supplies up to 50 A per unit
TH2883-5 impulse winding tester2Detects shorted turns and layer-to-layer insulation faults

Our team supported the acceptance with an on-site demonstration and technical guidance.

How to measure inductance under DC bias

  1. Connect the LCR meter to the bias current source, and the bias source to the test fixture, as shown in the manual.
  2. Perform open and short correction at the fixture. For the short correction, use a shorting plate, not clipped-together leads.
  3. Set the test frequency and AC level on the LCR meter.
  4. Set the bias current, or a list of current steps, on the bias source.
  5. Measure L and Q at each current step and compare them with the limits, for example the current at which L drops by a defined percentage.

Before you plan a similar set-up

  • Choose the bias set-up by the highest current you need. One TH1779A supplies up to 50 A; up to eight units in parallel supply up to 400 A.
  • High DC current heats the part and the fixture. Keep test times short and allow for the temperature rise when you compare results.

Frequently asked questions

Can I add DC bias to my existing Tonghui LCR meter? The TH1779A has a direct interface to compatible Tonghui LCR meters; the TH2840BL is designed for use with it. Ask us about your meter model.

How much bias current can I apply? Up to 50 A with one TH1779A, and up to 400 A with eight units in parallel.