TH2822 Series
Handheld Digital LCR Meters, up to 100 kHz
A compact, battery-powered LCR bridge family for evaluating capacitors, inductors, resistors, transformers and chip components at the bench, on the incoming-inspection line, or in the field. Four models share the same dual-terminal design and Auto LCR function, from a general-purpose base unit up to a 100 kHz / 0.1%-accuracy model with built-in DCR measurement.
Technical question about this series? Ask our team.
Key features
- Basic accuracy up to 0.1% (TH2822D/E), 0.25% on TH2822A/C
- Test frequency up to 100 kHz; selectable test level 0.3/0.6/1 Vrms on TH2822D/E
- 40,000-count primary reading, D/Q/θ resolution 0.0001
- Built-in DCR measurement on TH2822D/E, 1 Vdc test signal
- Auto LCR function identifies L, C or R and selects the matching equivalent circuit
- Dual compatible test terminals: 5-terminal test slot plus 3-terminal banana jacks
- Test speed up to 4 measurements/s (DCR 3/s), fast automatic ranging
- Percentage-display comparator with four tolerance bins: 1/5/10/20%
- Data hold, max/min/average recording, open/short zeroing
- Mini-USB interface, SCPI command set, free FastAccess PC software
4 models, one platform
TH2822A
Base model, 9 V battery
- Frequency
- 100 Hz – 10 kHz
- Basic accuracy
- 0.25%
- Test level
- 0.6 Vrms
- DCR
- —
TH2822C
Wider frequency range
- Frequency
- 100 Hz – 100 kHz
- Basic accuracy
- 0.25%
- Test level
- 0.6 Vrms
- DCR
- —
TH2822D
Multi-level test, built-in DCR
- Frequency
- 100 Hz – 10 kHz
- Basic accuracy
- 0.1%
- Test level
- 0.3/0.6/1 Vrms
- DCR
- Yes
TH2822E
Full-featured: 100 kHz + DCR
- Frequency
- 100 Hz – 100 kHz
- Basic accuracy
- 0.1%
- Test level
- 0.3/0.6/1 Vrms
- DCR
- Yes
TH2822A ships with a non-rechargeable 9 V battery; TH2822C/D/E use an 8.4 V NiMH rechargeable battery that keeps charging from the AC adapter whether the instrument is on or off. All models share the same 5-terminal/3-terminal test port, Auto LCR function and Mini-USB/SCPI interface.
Applications
Passive components
Capacitors, inductors, magnetic cores, resistors, piezoelectric devices, transformers, chip components and network components.
Other components
Printed circuit boards, relays, switches, cables and batteries.
Bench, line and field use
Battery-powered for incoming inspection and go/no-go sorting at the bench, or spot-checks on the shop floor and in the field.
Technical specifications
| Parameter | TH2822A | TH2822C | TH2822D | TH2822E |
|---|---|---|---|---|
| Display | LCD with backlight, primary + secondary dual display | |||
| Reading resolution | Primary parameter: 40,000 counts; secondary D/Q/θ: 0.0001 | |||
| Basic accuracy | 0.25% | 0.25% | 0.1% | 0.1% |
| Test frequency | 100 Hz – 10 kHz (4 points) | 100 Hz – 100 kHz (5 points) | 100 Hz – 10 kHz (4 points) | 100 Hz – 100 kHz (5 points) |
| Test signal level | 0.6 Vrms | 0.6 Vrms | 0.3/0.6/1 Vrms | 0.3/0.6/1 Vrms |
| DCR test level | — | — | 1 Vdc | 1 Vdc |
| Signal source impedance | 100 Ω | |||
| Test parameters | L, C, R, |Z|, D, Q, θ, ESR | L, C, R, |Z|, D, Q, θ, ESR | L, C, R, |Z|, DCR, D, Q, θ, ESR | L, C, R, |Z|, DCR, D, Q, θ, ESR |
| L range | 0.000 µH – 1000.0 H (res. 0.001 µH) | |||
| C range | 0.000 pF – 20.000 mF (res. 0.001 pF) | |||
| |Z|, R, ESR range | 0.0000 Ω – 10.000 MΩ (res. 0.0001 Ω) | |||
| DCR range | — | — | 0.0000 Ω – 20.000 MΩ (res. 0.0001 Ω) | 0.0000 Ω – 20.000 MΩ (res. 0.0001 Ω) |
| D range | 0.0000 – 9.999 (res. 0.0001) | |||
| Q range | 0.0000 – 9999 (res. 0.0001) | |||
| θ range | −179.99° – 179.99° (res. 0.01°) | |||
| Measurement speed, LCR | Fast 250 ms, slow 667 ms | |||
| Measurement speed, DCR | — | — | Fast 333 ms, slow 400 ms | Fast 333 ms, slow 400 ms |
| Equivalent circuit | Series, parallel | |||
| Ranging mode | Auto, manual | |||
| Correction | Open, short | |||
| Data hold and statistics | Data hold; max/min/average recording | |||
| Comparator | Percentage display, 4 tolerance bins: 1%, 5%, 10%, 20% | |||
| Test terminals | 3-terminal and 5-terminal, auto-switching | |||
| Interface | Mini-USB (virtual serial port), SCPI command set | |||
| Battery | 9 V alkaline (IEC 6LR61), non-rechargeable | 8.4 V NiMH, 200 mAh, rechargeable | 8.4 V NiMH, 200 mAh, rechargeable | 8.4 V NiMH, 200 mAh, rechargeable |
| Standby current | Max. 2 µA | Max. 18 µA | Max. 11 µA | Max. 11 µA |
| Battery life | Typ. 16 h (new alkaline) / typ. 6 h (full NiMH charge), backlight off | |||
| Charge time and current | — | Max. 160 min per cycle, max. 100 mA | Max. 160 min per cycle, max. 100 mA | Max. 160 min per cycle, max. 100 mA |
| AC power adapter | Input 220 V (±10%), 50 Hz (±5%); output 12–15 V DC | |||
| Auto power-off | 5/15/30/60 min or off, selectable (battery operation only); default 5 min | |||
| Operating environment | 0–40 °C, ≤ 90% R.H. | |||
| Weight / dimensions | 350 g; 190 (H) × 90 (W) × 41 (D) mm | |||
| Safety and EMC | IEC 61010-1:2001; IEC 61326-2-1:2005 | |||
Source: TH2822 Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- Mini-USBstandard (virtual serial port)
- SCPIstandard
- AC adapterstandard
Accessories
Standard accessories
| Item | Description |
|---|---|
| TH26028(GS) | EU-style AC power adapter |
| — | Mini-USB interface cable |
| TH26004F | Test leads, banana plug to alligator clip |
| TH26027AS | Four-terminal Kelvin test clip |
| TH26010B | Short-circuit calibration piece |
| TH26029C | SMD Kelvin test tweezers (TH2822C/D/E only, not included with TH2822A) |
Optional test fixtures
| Item | Description |
|---|---|
| TH26009C | SMD Kelvin test tweezers |
More fixtures are available on request from Techmize or your authorized distributor.
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH2822 Series?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.