Low-Current & Insulation Resistance

TH2690 Series

Femtoammeters, Picoammeters, Electrometers and High-Resistance Meters

Eight models built around a common 6½-digit measurement platform, for the weak-current and high-resistance measurements that general-purpose digital multimeters cannot reach. The electrometer / high-resistance family (TH2690, TH2690A, TH2690H, TH2695, TH2695A) resolves current down to 0.01 fA and measures resistance up to 3 EΩ (10¹⁸ Ω) using a built-in voltage source of up to ±3000 V. The compact TH2691 / TH2691A / TH2691H family offers the same femtoamp-level ammeter core for current-only applications at lower cost.

TH2690TH2690ATH2690HTH2691TH2691ATH2691HTH2695TH2695A CE

Technical question about this series? Ask our team.

Key features

  • Current resolution down to 0.01 fA (10⁻¹⁷ A)
  • Resistance measurement up to 3 EΩ (10¹⁸ Ω) via built-in ±3000 / ±1500 / ±1000 V source
  • 5.0" capacitive touch colour LCD, 6½-digit measurement resolution
  • Four measurement modes: high-resistance meter, voltmeter, ammeter, electrometer
  • Independent current and voltage measurement inputs
  • Data reading rate up to 5000 readings/second, with time-domain waveform capture
  • Charge measurement down to a 2 nC range (1 fC resolution)
  • Programmable voltage output: DC, single/double-step scan, list scan, square wave/pulse
  • Dedicated temperature and humidity probe input; sorting comparator with up to 7 limits
  • SCPI / MODBUS standard; RS232, USB Host/Device, LAN, HANDLER, INTERLOCK interfaces

8 models, one platform

ModelFunctionsCurrent resolutionLowest current rangeMax. resistanceVoltage sourceVoltage measurementChargeQuote
TH2690 R / V / I / Q0.1 fA20 pA100 PΩ±1000 V1 µV – 20 V1 fC – 2 µC Quote
TH2690A R / V / I1 fA2 nA1 PΩ±1000 V1 µV – 20 V— Quote
TH2690H R / V / I / Q0.01 fA2 pA1000 PΩ±1000 V1 µV – 20 V1 fC – 2 µC Quote
TH2691 I0.1 fA20 pA———— Quote
TH2691A I1 fA2 nA———— Quote
TH2691H I0.01 fA2 pA———— Quote
TH2695 R / V / I / Q0.01 fA2 pA3000 PΩ (3 EΩ)±3000 V30 mV – 3000 V1 fC – 2 µC Quote
TH2695A R / V / I / Q0.01 fA2 pA1500 PΩ±1500 V15 mV – 1500 V1 fC – 2 µC Quote

R = high-resistance meter · V = voltmeter · I = ammeter · Q = electrometer (charge). All eight models share the 20 mA top current range and 6½-digit resolution; the lowest current range depends on the model. TH2691 / TH2691A / TH2691H are compact ammeter-only versions without voltage source, voltage input, resistance, charge or temperature/humidity input.

Current measurement ranges

RangeAccuracyResolutionApplicable models
2 pA±(1% + 3 fA)0.01 fATH2695 / TH2695A / TH2690H / TH2691H
20 pA±(0.5% + 3 fA)0.1 fATH2695 / TH2695A / TH2690 / TH2690H / TH2691 / TH2691H
200 pA±(0.5% + 5 fA)0.1 fATH2695 / TH2695A / TH2690 / TH2690H / TH2691 / TH2691H
2 nA±(0.2% + 50 fA)1 fAAll models
20 nA±(0.2% + 3 pA)10 fAAll models
200 nA±(0.2% + 5 pA)100 fAAll models
2 µA±(0.1% + 50 pA)1 pAAll models
20 µA±(0.05% + 500 pA)10 pAAll models
200 µA±(0.05% + 5 nA)100 pAAll models
2 mA±(0.05% + 50 nA)1 nAAll models
20 mA±(0.05% + 500 nA)10 nAAll models

Applications

Material characterization

Semiconductors, nanomaterials, polymers, dielectric and ferroelectric materials, graphene, ceramics, biological materials, rubber, thin films, metals and organic materials.

Component leakage and insulation

Leakage current and insulation-resistance testing of capacitors, resistors, diodes, transistors, sensors, TFT/CNT devices, opto-electronic and nano-devices, solar cells, switches and relays.

Resistivity and I–V testing

Volume/surface resistivity with the optional TH26090 test fixture; third-generation semiconductor I–V characterization; fast charge testing of capacitive devices.

Industries and application examples

Technical specifications

ParameterTH2690TH2690ATH2690HTH2691TH2691ATH2691HTH2695TH2695A
Display 5.0" capacitive touch colour LCD, 6½-digit measurement resolution
Resistance 1 MΩ / 10 MΩ range ±(0.135% + 1 Ω) / ±(0.135% + 10 Ω)
Resistance 100 MΩ / 1 GΩ range ±(0.185% + 100 Ω) / ±(0.285% + 1 kΩ)
Resistance 10 GΩ / 100 GΩ range ±(0.41% + 10 kΩ) / ±(0.41% + 100 kΩ)
Resistance 1 TΩ range ±(0.45% + 1 MΩ)
Resistance 10 TΩ / 100 TΩ range ±(0.75% + 10 MΩ) / ±(2.6% + 100 MΩ); not on TH2690A
Resistance measurement models TH2695, TH2695A, TH2690, TH2690A, TH2690H (not TH2691 / TH2691A / TH2691H)
Max. measurable resistance TH2695 3000 PΩ (3 EΩ) · TH2695A 1500 PΩ · TH2690H 1000 PΩ · TH2690 100 PΩ · TH2690A 1000 TΩ (1 PΩ)
Voltage measurement, TH2690 / TH2690A / TH2690H 2 V range ±(0.05% + 40 µV), 1 µV res. · 20 V range ±(0.05% + 400 µV), 10 µV res. · input impedance > 200 TΩ
Voltage measurement, TH2695 300 V range ±(0.05% + 600 µV), 100 µV res. · 3000 V range ±(0.05% + 6 mV), 1 mV res. · input impedance 100 MΩ
Voltage measurement, TH2695A 150 V range ±(0.05% + 300 µV), 100 µV res. · 1500 V range ±(0.05% + 300 µV), 1 mV res. · input impedance 50 MΩ
Charge measurement 2 nC ±(0.5% + 50 fC), 1 fC res. · 20 nC ±(0.5% + 500 fC) · 200 nC ±(0.5% + 5 pC) · 2 µC ±(0.5% + 50 pC); TH2695 / TH2695A / TH2690 / TH2690H
Voltage source, TH2690 / TH2690A / TH2690H 20 V ±(0.05% + 2 mV), ±20 mA · 1000 V ±(0.05% + 100 mV), ±1 mA; 1 mV resolution
Voltage source, TH2695 40 V ±(0.05% + 4 mV), ±20 mA · 400 V ±(0.05% + 40 mV), ±10 mA · 3000 V ±(0.05% + 300 mV), ±10 mA; 1 mV resolution
Voltage source, TH2695A 20 V ±(0.05% + 2 mV), ±20 mA · 200 V ±(0.05% + 20 mV), ±10 mA · 1500 V ±(0.05% + 150 mV), ±10 mA; 1 mV resolution
Source functions DC, scan (single-step, double-step, list scan), ARB (square wave with pulse output, 10 ms min. pulse width, 1 ms resolution)
Temperature probe input −40 to −10 °C ±1 °C · 10 to 55 °C ±0.5 °C · 55 to 80 °C ±1 °C
Humidity probe input 0–20% RH ±4% · 20–80% RH ±3% · 80–100% RH ±4%
View modes Meter, graph (I‑t, V‑t, R‑t, Q‑t, I‑R), histogram, scroll
Input terminals Current input triaxial BNC; voltage input triaxial BNC (high-voltage BNC on TH2695 / TH2695A); GROUND banana jack
Power supply AC 90–264 V, 50/60 Hz or DC 127–370 V · 60 W
Operating / storage environment 0–45 °C, 30–80% RH / −20–60 °C, 10–90% RH (no condensation)
Accuracy-rated environment 23 °C ± 5 °C, 30–80% RH · warm-up 1 hour · calibration cycle 1 year
Dimensions / weight, TH2690/A/H and TH2691/A/H Rack mount 215 × 88 × 335 mm · Bench 235 × 111 × 440 mm · 3.5 kg
Dimensions / weight, TH2695A Rack mount 215 × 88 × 335 mm · Bench 235 × 111 × 440 mm · 4 kg
Dimensions / weight, TH2695 Rack mount 215 × 88 × 480 mm · Bench 235 × 111 × 590 mm · 5.5 kg

Source: TH2690 Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.

Interfaces

  • RS232standard
  • USB Hoststandard
  • USB Devicestandard
  • LANstandard
  • HANDLERstandard
  • INTERLOCKstandard (not TH2691/A/H)
  • SCPI / MODBUSstandard
  • Analog / D-A out±2 VFS

Accessories

Standard accessories

ItemDescription
—EU power cable (Schuko, CEE 7/7)
TH26017USB interface cable
TH26058Fφ4 alligator clip (red/black)
TH26058DConnecting cable
TH2690_THSTemperature and humidity probe (not TH2691/A/H)

Optional test fixtures

ItemDescription
TH26090Resistivity test fixture (Φ26/38/50 mm electrodes)
TH26076BShielding box
TH26058BTriaxial-to-alligator-clip cable (1.5 m)
TH26058SHigh-voltage BNC RF test cable (TH2695 / TH2695A)

More fixtures and adapters are available on request from Techmize or your authorized distributor.

Photos of fixtures and accessories: click an item code above, or browse all accessories.

Downloads

Interested in the TH2690 Series?

Send us the model, quantity and your application. Our team in Ismaning will answer by email.

Request a quote