Digital Multimeters & Data Acquisition
TH2554 Series
Modular Data Acquisition / Switch System
A high-precision, high-stability, multi-channel scanning data acquisition system built around a 5-slot modular mainframe. The TH2554 combines a 6½-digit DMM measurement module with plug-in multiplexer modules to support up to 160 switched channels, for mixed-signal data logging, production testing and R&D bench use.
Technical question about this series? Ask our team.
Key features
- 7-inch capacitive touchscreen, 800 × 480, menu-driven operation
- 6½-digit DMM module with DCV, DCI, ACV, ACI, 2-wire/4-wire resistance, frequency, period, temperature, capacitance, continuity, diode and ratio
- 5-slot modular mainframe with up to 160 switched channels and automatic module recognition
- Three multiplexer modules, including the TH2554-03 high-speed module (solid-state relays, 300 CH/s)
- Reading rate up to 1,000 readings/second; scan buffer of up to 100,000 timestamped readings
- Independent per-channel configuration: function, Mx+b scaling, pass/fail limits
- Built-in cold-junction compensation on the multiplexer modules for thermocouple measurement
- Direct USB data logging in CSV, TXT or DAT format
- Standard SCPI; RS232, USB Host, USB Device and LAN standard; GPIB or Handler optional
Model
TH2554
5-slot data acquisition / switch mainframe
- Module slots
- 5
- Max. channels
- 160
- Reading rate
- Up to 1,000 readings/s
- Scan buffer
- 100,000 readings
The TH2554-DMM measurement module is mandatory. Any combination of up to 5 modules can be fitted, with automatic module recognition by the mainframe. All multiplexer modules provide built-in cold-junction compensation (CJC) for thermocouple measurement. The TH2554-03 measures voltage only (no current channels).
Plug-in modules
| Feature | TH2554-DMM | TH2554-01 | TH2554-02 | TH2554-03 |
|---|---|---|---|---|
| Type | Measurement module (mandatory) | 20-channel differential multiplexer | 40-channel differential multiplexer | 20-channel high-speed multiplexer |
| Channels | — | 20 × 2-wire / 10 × 4-wire + 2 current | 40 × 2-wire / 20 × 4-wire + 2 current | 20 × 2-wire / 10 × 4-wire |
| Resolution | 6½ digits | — | — | — |
| Scan speed | — | 90 CH/s | 90 CH/s | 300 CH/s |
| Max. rating | 300 V / 3 A (input) | 300 V / 1 A | 300 V / 1 A | 60 V (voltage only) |
| Relay type | — | Electromechanical | Electromechanical | Solid-state |
| Cold-junction compensation | — | ✓ | ✓ | ✓ |
Applications
Industrial
Temperature testing for motors, transformers and magnetic cores, thermal-rise evaluation, sintering furnaces, new-energy batteries, charging stations, automotive motors, LED lamps and semiconductor chips.
Agriculture and environmental
Temperature monitoring in greenhouses, seed and cold-storage facilities and outdoor meteorological observation sites.
Chemical, medical and transportation
Temperature monitoring for reaction furnaces and production equipment, hospital wards and pharmaceutical storage, refrigerated transportation and automotive electronics; also livestock sheds, cold-storage warehouses and traffic/expressway infrastructure monitoring.
Technical specifications
| Parameter | TH2554 |
|---|---|
| DC voltage | Ranges 100.0000 mV – 300.000 V, resolution down to 0.1 µV; 1-year accuracy on 10 V range ±(0.0035 + 0.0005) |
| DC current | Ranges 10 µA – 3 A, resolution down to 0.01 nA; 1-year accuracy on 1 mA range ±(0.050 + 0.006) |
| DC resistance (2W / 4W) | Ranges 10.00000 Ω – 100.0000 MΩ; measurement range 10 µΩ – 120 MΩ; 1-year accuracy on 1 kΩ – 100 kΩ ranges ±(0.0100 + 0.0010) |
| AC voltage | 0.1 µV – 300 V, 3 Hz – 300 kHz; 1-year accuracy 10 Hz – 20 kHz ±(0.06 + 0.03) |
| AC current | 1 nA – 3 A; 1-year accuracy 100 µA – 100 mA, 3 Hz – 5 kHz ±(0.10 + 0.04) |
| Capacitance | 0.0001 nF – 10.000 mF; 1-year accuracy 0.2 + 0.05 to 1.0 + 0.50 depending on range |
| Frequency / period | 3 Hz – 300 kHz |
| Temperature sensors | Thermocouple (B/E/J/K/N/R/S/T), thermistor, RTD (2W/3W/4W, PT100/PT385) |
| Continuity / diode | 1 kΩ range ±(0.010 + 0.030) / 5 V range, 1 mA ±(0.010 + 0.030), 1-year |
| Accuracy conditions | 60-minute warm-up, 1-year calibration cycle, TCAL ± 5 °C; ±(% of reading + % of range) |
| Scan buffer | Up to 100,000 timestamped readings |
| Reading rate | Up to 1,000 readings/second |
| Display | 7" capacitive colour touchscreen, 800 × 480 |
| Remote command set | SCPI |
| Power supply | 90 – 264 V AC, 50/60 Hz (1 ± 5%), < 250 VA |
| Operating environment | 0 – 40 °C, ≤ 90% RH |
| Safety / EMC | Class I; Directive 2014/35/EU, EN 61010-1:2010+A1:2019; Directive 2014/30/EU, EN 61326-1:2021 |
| Dimensions (W × H × D) | 235 × 154 × 480 mm |
| Weight | 7.8 kg |
Source: TH2554 Series Datasheet, Rev. 2026-11. Specifications subject to change without notice.
Interfaces
- RS232standard
- USB Hoststandard
- USB Devicestandard
- LANstandard
- SCPIstandard
- IEEE-488 / GPIBoptional (shares slot with Handler)
- Handleroptional (shares slot with GPIB)
Accessories
Standard accessories
| Item | Description |
|---|---|
| — | EU power cable (Schuko, CEE 7/7) |
| — | 1 A fuse × 2 |
| — | Operation manual, test report, warranty card |
Optional test fixtures
| Item | Description |
|---|---|
| TH2554-DMM | 6½-digit DMM measurement module (mandatory) |
| TH2554-01 | 20-channel differential multiplexer module |
| TH2554-02 | 40-channel differential multiplexer module |
| TH2554-03 | 20-channel high-speed multiplexer module |
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH2554 Series?